TY - JOUR
T1 - δ(18O/16O) Determinations in Water Using Inductively Coupled Plasma–Tandem Mass Spectrometry
AU - Lancaster, Shaun
AU - Irrgeher, Johanna
AU - Dallmayr, Remi
AU - Conrad, Elisa
AU - Hörhold, Maria
AU - Bohleber, Pascal
AU - Behrens, Melanie
AU - Camin, Federica
AU - Žagar, Klara
AU - Vreča, Polona
AU - Prohaska, Thomas
N1 - Publisher Copyright: © 2025 The Authors. Published by American Chemical Society. This article is licensed under CC-BY-NC-ND 4.0
PY - 2025/9/16
Y1 - 2025/9/16
N2 - A fully validated measurement approach for the determination of oxygen isotope ratios in water using inductively coupled plasma–tandem mass spectrometry (ICP-MS/MS) is described for the first time. Deuterium was used as a reaction gas to mass-shift the oxygen isotopes to the OD3+product ion, allowing for the removal of polyatomic ions (16O1H+and16O1H2+) generated in the plasma that interfere on the less abundant oxygen stable isotopes (17O and18O). The developed methodology was validated for δVSMOW-SLAP(18O/16O) determinations using four internationally recognized IAEA reference materials, as well as 11 in-house laboratory isotope reference materials that were characterized by isotope ratio mass spectrometry and cavity ring-down spectroscopy. δVSMOW-SLAP(18O/16O) values determined by ICP-MS/MS and the reference methods overlapped within the uncertainties and showed no significant difference for water materials with a low matrix load. Elevated levels of sodium, chloride, and silicon in the matrix lead to isotope ratio shifts of up to 12‰ in both the positive and negative directions. Additionally, comparisons with indicative values from the reference materials demonstrated that δVSMOW-SLAP(17O/16O) determinations are also possible using the ICP-MS/MS approach. While the uncertainty of the developed approach (median uncertainty uc(δVSMOW-SLAP(18O/16O)) = 0.70‰) is not currently able to match that of existing methodologies, it is envisioned that oxygen isotope ratio determinations can be performed alongside multielemental analysis by a wider community for applications where lower uncertainties are not required.
AB - A fully validated measurement approach for the determination of oxygen isotope ratios in water using inductively coupled plasma–tandem mass spectrometry (ICP-MS/MS) is described for the first time. Deuterium was used as a reaction gas to mass-shift the oxygen isotopes to the OD3+product ion, allowing for the removal of polyatomic ions (16O1H+and16O1H2+) generated in the plasma that interfere on the less abundant oxygen stable isotopes (17O and18O). The developed methodology was validated for δVSMOW-SLAP(18O/16O) determinations using four internationally recognized IAEA reference materials, as well as 11 in-house laboratory isotope reference materials that were characterized by isotope ratio mass spectrometry and cavity ring-down spectroscopy. δVSMOW-SLAP(18O/16O) values determined by ICP-MS/MS and the reference methods overlapped within the uncertainties and showed no significant difference for water materials with a low matrix load. Elevated levels of sodium, chloride, and silicon in the matrix lead to isotope ratio shifts of up to 12‰ in both the positive and negative directions. Additionally, comparisons with indicative values from the reference materials demonstrated that δVSMOW-SLAP(17O/16O) determinations are also possible using the ICP-MS/MS approach. While the uncertainty of the developed approach (median uncertainty uc(δVSMOW-SLAP(18O/16O)) = 0.70‰) is not currently able to match that of existing methodologies, it is envisioned that oxygen isotope ratio determinations can be performed alongside multielemental analysis by a wider community for applications where lower uncertainties are not required.
UR - http://www.scopus.com/inward/record.url?scp=105017377494&partnerID=8YFLogxK
U2 - 10.1021/acs.analchem.5c02607
DO - 10.1021/acs.analchem.5c02607
M3 - Article
AN - SCOPUS:105017377494
SN - 0003-2700
VL - 97
SP - 20788
EP - 20797
JO - Analytical chemistry
JF - Analytical chemistry
IS - 38
ER -