Zur Hauptnavigation wechseln Zur Suche wechseln Zum Hauptinhalt wechseln

Experimental results on an optical pH measurement system for bioreactors

  • Andrej Holobar
  • , Bernhard H. Weigl
  • , Wolfgang Trettnak
  • , Roman Benes
  • , Hartmut Lehmann
  • , Nena V. Rodriguez
  • , Arnold Wollschlager
  • , Paul O'Leary
  • , Peter Raspor
  • , Otto S. Wolfbeis
  • Joanneum Research Forschungsgesellschaft mbH, Graz
  • Technische Universität Graz
  • Slovak Academy of Sciences, Bratislava
  • Universität Laibach

Publikation: Beitrag in FachzeitschriftArtikelForschungBegutachtung

30 Zitate (Scopus)

Abstract

Properties of a pH-sensitive membrane have been tested with respect to its possible application in bioreactors. The influences of ionic strength, buffer type, buffer capacity, color and turbidity of the solution on the response of the membrane are studied. The developed membrane shows high reproducibility and long-term stability. The pH-sensing membrane can be sterilized chemically and by using heat. With the present opto-electronic set-up, an accuracy of ±0.1 pH between pH 6 and 10 can be achieved.
OriginalspracheEnglisch
Seiten (von - bis)425-430
Seitenumfang6
FachzeitschriftSensors and Actuators: B. Chemical
Jahrgang11.1993
Ausgabenummer1-3
DOIs
PublikationsstatusVeröffentlicht - 1 März 1993
Extern publiziertJa

Dieses zitieren